Affiliation of Author(s):航空学院
Journal:Zhendong Ceshi Yu Zhenduan
Abstract:Piezoelectric actuators have been widely applied in scanning tunneling microscope (STM). A controller combining inverse hysteresis model and PID control law was designed for the STM to compensate for the hysteresis nonlinearity of piezoelectric actuators and to improve the scanning accuracy. For the purpose of comparing the proposed controller with classical displacement feedback control law, experimental system was built in this investigation. The relative measuring error of grating period and width of grating line were 4.41% and 2.65% using the classic displacement feedback control method respectively. And they were reduced to 1.26% and 0.27% in proposed method. Experimental results indicate that the error caused by hysteresis nonlinearity has been well compensated, and the scanning accuracy is improved with the help of such controller. © 2017, Editorial Department of JVMD. All right reserved.
ISSN No.:1004-6801
Translation or Not:no
Date of Publication:2017-04-01
Co-author:孙宏君,Jinhao Qiu,Chen, Yuansheng,Kongjun Zhu
Correspondence Author:jihongli
Date of Publication:2017-04-01
季宏丽
+
Gender:Female
Education Level:南京航空航天大学
Alma Mater:南京航空航天大学
Paper Publications
Control of Hysteresis Nonlinearity for Scanning Stage in Scanning Tunneling Microscope
Date of Publication:2017-04-01 Hits: